• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 12FR 43 IRIS FEEDING TUBE ENF; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 12FR 43 IRIS FEEDING TUBE ENF; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 461243E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
 
Event or Problem Description
The customer reported that the tube broke at y junction.There was leaking observed when the feeding tube flushed after giving medication and with tube feeding running.No patient harm reported other than discomfort and inconvenience of replacing the feeding tube earlier than necessary.The feeding tube was placed on (b)(6) 2021.
 
Additional Manufacturer Narrative
Section d4: lot number field has been updated, based on the additional information provided by the customer.
 
Additional Manufacturer Narrative
A device history record review could not be performed because the serial number was not received with the complaint.The device was received for evaluation and the reported condition has been confirmed.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12FR 43 IRIS FEEDING TUBE ENF
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
15 hampshire street
mansfield, MA 02048
5086183640
MDR Report Key11553396
Report Number1282497-2021-09988
Device Sequence Number4204232
Product Code KNT
UDI-Device Identifier20884521742229
UDI-Public20884521742229
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number461243E
Device Catalogue Number461243E
Device Lot Number1255050120
Was Device Available for Evaluation? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer03/16/2021
03/16/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date03/26/2021
03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-