• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US APG+ SIZER PIN GUIDE 44+0; EXTREMITY INSTRUMENTS : PIN GUIDES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US APG+ SIZER PIN GUIDE 44+0; EXTREMITY INSTRUMENTS : PIN GUIDES Back to Search Results
Model Number 2236-00-002
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during total shoulder it was noted that the 40 and 44 sizers were not holding on the guide due to wear.A second set was opened and these were also worn.This caused a 5-minute delay to surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned, thus the reported event could not be confirmed.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : no lot information available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported wear and inability to retain the handle guide.The investigation did not establish that a broader investigation or corrective action was necessary.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APG+ SIZER PIN GUIDE 44+0
Common Device Name
EXTREMITY INSTRUMENTS : PIN GUIDES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11553529
Report Number1818910-2021-05806
Device Sequence Number17165818
Product Code HWT
UDI-Device Identifier10603295105695
UDI-Public10603295105695
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2236-00-002
Device Catalogue Number223600002
Device Lot NumberPG233151
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Initial Date Received by Manufacturer 03/09/2021
Supplement Date Received by Manufacturer04/23/2021
04/27/2021
08/24/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date04/25/2021
05/11/2021
08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-