BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 4/150/130; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
383281 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor angiographic material was returned for technical investigation.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.The instrument was delivered in a leak-proof and pressure-tight condition.Based on the conducted investigations no manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The passeo-35 balloon catheter was selected for the treatment of a calcified lesion in the sfa.Although the doctor tried for several times, the balloon could not be inflated.
|
| |
|