• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ELEGANCE; COMPRESSOR, AIR, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ELEGANCE; COMPRESSOR, AIR, PORTABLE Back to Search Results
Model Number 1102460
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received an innospire elegance device for evaluation with the complaint that the device would not switch on.There was no patient harm or injury reported.During the evaluation of the device at the manufacturer's service center, an issue related to the device's ac power cord was observed.There was evidence of exposure to the conducting wires.The manufacturer concludes that this exposure was caused by the ac power cord being exposed to forces beyond the intended design.The innospire elegance device is intended to provide a source of compressed air for medical purposes.It is to be used with a pneumatic (jet) nebulizer to produce aerosol particles of medication for respiratory therapy for children and adults.Labeling warns the "it is recommended to have a backup device for respiratory delivery in case a situation arises when your nebulizer can not be used".This device meets all relevant ul and iec standards.This device is not labeled for life support.Based on the information available, the manufacturer concludes no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ELEGANCE
Common Device Name
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, PO20 2FT
UK  PO20 2FT
MDR Report Key11554595
Report Number9681154-2021-00008
Device Sequence Number19845469
Product Code BTI
UDI-Device Identifier00383730001753
UDI-Public00383730001753
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K042655
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1102460
Device Catalogue Number1102460
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2021
Initial Report FDA Received Date03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-