• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE S MRI SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE S MRI SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 693558
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dyspnea (1816); Fever (1858); Unspecified Infection (1930); Chills (2191)
Date of Event 06/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient developed bacteremia.The patient presented to the emergency department with worsening shortness of breath, cough, chills, and a fever.It was noted that the patient was symptomatic with similar symptoms a month prior and tested negative for coronavirus 2019.A transthoracic echocardiogram (tte) revealed a mobile mass in the right ventricle attached to the right ventricular (rv) lead that is consistent with vegetation.Peripheral blood cultures collected on admission grew enterococcus faecalis and aerobic bottle gram positive cocci in chains.The cardiac resynchronization therapy defibrillator (crt-d) system was removed, and a peripherally inserted central catheter (picc) line was placed for continued antibiotics treatment.Follow up blood cultures were negative, and a deep vein thrombosis (dvt) was noticed in the left popliteal.The patient is a participant in a clinical study.No further patient complications have been reported as a result of this event.
 
Event or Problem Description
Additional information received indicated the dvt was identified via bilateral lower extremity ultrasound as non-occlusive left popliteal dvt.The patient was started on anticoagulation therapy, however it was stopped due to the patient history of severe epitaxies while on anticoagulation therapy.A inferior vena cava filter was placed without complication and the dvt was indicated as resolved.It was also reported that the dvt was not related to the product.
 
Additional Manufacturer Narrative
Product event summary the full lead was returned and analyzed.No anomalies were found.Visual analysis of the lead indicated apparent explant damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11554918
Report Number2649622-2021-05543
Device Sequence Number3845038
Product Code LWS
UDI-Device Identifier00643169535909
UDI-Public00643169535909
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/14/2018
Device Model Number693558
Device Catalogue Number693558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer03/26/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date04/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTBA1Q1 CRTD, 383059 LEAD, 459878 LEAD; DTBA1Q1 CRTD, 383059 LEAD, 459878 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient Weight88
-
-