• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PIN TRIAL SHELL 60 STD PROFILE; HIP INSTRUMENTS : ACETABULAR TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PIN TRIAL SHELL 60 STD PROFILE; HIP INSTRUMENTS : ACETABULAR TRIALS Back to Search Results
Model Number 2217-01-060
Device Problems Corroded (1131); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Complaint from the sterile department due to corrosion on ce marking or numbers.Looks like corrosion on markings, brownish color! no new procedures in sterile depart, no new chemicals.
 
Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument cannot confirm the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN TRIAL SHELL 60 STD PROFILE
Type of Device
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11556686
MDR Text Key246848574
Report Number1818910-2021-05881
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295097716
UDI-Public10603295097716
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2217-01-060
Device Catalogue Number221701060
Device Lot NumberSO2039485
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2021
Date Manufacturer Received04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-