MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL G2 CURVED JAW 14CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL G2 CURVED JAW 14CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLG2C14
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: t94z57.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2c14 device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with opening and closing of the jaws.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.Please reference the instruction for use for more information.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure the jaw of the enseal could not open near the end of the surgery.They cleaned the jaw and the jaw still could not open.It was chosen to not continue the surgery with enseal.Instead, she used hand sewed.The procedure was completed successfully.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 CURVED JAW 14CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key11557887
Report Number3005075853-2021-01555
Device Sequence Number4088278
Product Code GEI
UDI-Device Identifier10705036014164
UDI-Public10705036014164
Combination Product (Y/N)N
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberNSLG2C14
Device Catalogue NumberNSLG2C14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2021
Initial Report FDA Received Date03/23/2021
Patient Sequence Number1
-
-