NIPRO CORPORATION (ODT) NIPRO CELLENTIA-H; DIALYZER Back to Search Results
Model Number DD+CT19H
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
1 patient/ 3 occurrences with the same lot # reported.Patient lost blood with the 1st occurrence.The 2nd and 3rd occurrence occurred prior to initiating treatment, the machine kept alarming for air.On the 3rd occurrence staff changed the dialyzer with a different lot # and treatment was completed without complications.Additional information received 03/02/2021: priming conditions - they prime with a bfr of 100 and flushed the dialyzer with at least a l of saline.Dialysis conditions - bfr 400, dfr - 800, treatment time 4 hrs and no heparin.Brand/manufacturer name & models of devices used together with dialyzer - gambro machine and lines.How much blood did the patient lost due to changing the dialyzer - 300 ml.Was any medical actions taken- none as their hgb (as claimed when they rechecked) was stable.Where did the bubbles generate - middle to top portion of the dialyzer.
 
Event or Problem Description
1 patient/ 3 occurrences with the same lot # reported.Patient lost blood with the 1st occurrence.The 2nd and 3rd occurrence occurred prior to initiating treatment, the machine kept alarming for air.On the 3rd occurrence staff changed the dialyzer with a different lot # and treatment was completed without complications.Additional information received 03/02/2021: 1.Priming conditions - they prime with a bfr of 100 and flushed the dialyzer with at least a l of saline.2.Dialysis conditions - bfr 400, dfr - 800, treatment time 4 hrs and no heparin.3.Brand/manufacturer name & models of devices used together with dialyzer - gambro machine and lines.4.How much blood did the patient lost due to changing the dialyzer - 300 ml.5.Was any medical actions taken- none as their hgb (as claimed when they rechecked) was stable.6.Where did the bubbles generate - middle to top portion of the dialyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO CELLENTIA-H
Common Device Name
DIALYZER
Manufacturer (Section D)
NIPRO CORPORATION (ODT)
8-7, hanuki-yachi, niida-aza
ohdate-shi, akita 018-5 794
JA  018-5794
MDR Report Key11558883
Report Number9610987-2021-00002
Device Sequence Number19957548
Product Code KDI
UDI-Device Identifier40383790004402
UDI-Public40383790004402
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K160444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberDD+CT19H
Device Lot Number20H03D
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer02/25/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-