| Model Number |
DD+CT19H |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
1 patient/ 3 occurrences with the same lot # reported.Patient lost blood with the 1st occurrence.The 2nd and 3rd occurrence occurred prior to initiating treatment, the machine kept alarming for air.On the 3rd occurrence staff changed the dialyzer with a different lot # and treatment was completed without complications.Additional information received 03/02/2021: priming conditions - they prime with a bfr of 100 and flushed the dialyzer with at least a l of saline.Dialysis conditions - bfr 400, dfr - 800, treatment time 4 hrs and no heparin.Brand/manufacturer name & models of devices used together with dialyzer - gambro machine and lines.How much blood did the patient lost due to changing the dialyzer - 300 ml.Was any medical actions taken- none as their hgb (as claimed when they rechecked) was stable.Where did the bubbles generate - middle to top portion of the dialyzer.
|
| |
|
Event or Problem Description
|
|
1 patient/ 3 occurrences with the same lot # reported.Patient lost blood with the 1st occurrence.The 2nd and 3rd occurrence occurred prior to initiating treatment, the machine kept alarming for air.On the 3rd occurrence staff changed the dialyzer with a different lot # and treatment was completed without complications.Additional information received 03/02/2021: 1.Priming conditions - they prime with a bfr of 100 and flushed the dialyzer with at least a l of saline.2.Dialysis conditions - bfr 400, dfr - 800, treatment time 4 hrs and no heparin.3.Brand/manufacturer name & models of devices used together with dialyzer - gambro machine and lines.4.How much blood did the patient lost due to changing the dialyzer - 300 ml.5.Was any medical actions taken- none as their hgb (as claimed when they rechecked) was stable.6.Where did the bubbles generate - middle to top portion of the dialyzer.
|
| |
|
Search Alerts/Recalls
|