COVIDIEN MFG SOLUTIONS S.A. MAHURKAR ELITE; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number 8888213216
Medical Device Problem Codes Material Frayed (1262); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, an attempt was made to insert the catheter with the guidewire, the procedure was performed as usual and when removing the guidewire after post-implantation of the catheter, it did not come out.Marked resistance was observed which forced the catheter and guidewire assembly to be withdrawn completely.After the removal, an unravelling of the guidewire was observed at the tip of the catheter causing the integrity of the needle to be lost and 70% of it comes out of the lumen.The catheter was not repaired, there was no leak, tego was not utilized, there was no luer adapter issue and there was no patient symptoms or complications associated with this event.There was no medical or surgical intervention needed to prevent a permanent impairment of a function and the event did not lead to or extend patient hospitalization.There was no reported patient injury.
 
Additional Manufacturer Narrative
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection of the image depicted the green tip of the catheter cannula, and the guide wire was unraveled.It was reported that the guide wire was unable to be withdrawn, and the guide wire was unraveled.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR ELITE
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
MDR Report Key11558935
Report Number3009211636-2021-00069
Device Sequence Number4845640
Product Code MPB
UDI-Device Identifier10884521125636
UDI-Public10884521125636
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K120674
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2024
Device Model Number8888213216
Device Catalogue Number8888213216
Device Lot Number1925300147
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2021
Supplement Date Received by Manufacturer05/10/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date06/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
-
-