No samples or photos were received for analysis and investigation; for that reason, the malfunction reported by the customer could not be confirmed.In addition, a complaint search for lot 0k04372 and malfunction code skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (moldable only) was carried out and as a result, no additional complaint was found; therefore, no trend is observed.As per complaint manufacturing investigation procedure work instructions (wi), it is not required to open a nonconformance report (ncr) for type 2 complaints which were not confirmed.Lot 0k04372 was manufactured on 11/05/2020, ats#2, with a total of (b)(4) units.Complaint investigator performed a batch record review on 04/30/2021, description natura stoma flatmold 33/57mm1x10pkgb/nl to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct.System application product (sap) material 1222851 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|