SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568S
Medical Device Problem Codes Overheating of Device (1437); Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during set up or inspection for an unknown surgery the light source 500xl xenon was showing a e12 error when turned on, it was unable to get any light.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.The complaint of e12 error and overheating was confirmed.The complaint investigation has concluded that the cause of the overheating and e12 errors is a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL LIGHT SOURCE 500XL XENON
Common Device Name
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11560993
Report Number1643264-2021-01128
Device Sequence Number11957475
Product Code FFS
UDI-Device Identifier03596010650504
UDI-Public03596010650504
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K994084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200568S
Device Catalogue Number72200568S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2021
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer04/22/2021
Initial Report FDA Received Date03/24/2021
Supplement Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-