| Model Number |
SSLP-265 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
01/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: opening curved on radius and weight to the spec.A visual and microscopic analysis was performed which identified: striated opening on radius and wear abrasion.Base on the device analysis the final assessment is: a striated opening on radius assessed as surgical damage consist in the use of some surgical tool.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.
|
| |
|