Model Number 497-28-105
Health Effect - Clinical Code Discomfort (2330)
Date of Event 02/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to patient discomfort.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as discomfort.The previous surgery and the surgery detailed in this event occurred 24 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to discomfort.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, improper surgical technique, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS AND SLR-PLUS
Common Device Name
HEAD, FEMORAL, COCR 28/+10.5MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin 78758-5445
MDR Report Key11562362
Report Number1644408-2021-00181
Device Sequence Number14527209
Product Code JDI
UDI-Device Identifier00888912087285
UDI-Public(01)00888912087285
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2021
Device Model Number497-28-105
Device Catalogue Number497-28-105
Device Lot Number632C1128
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/24/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/24/2021
Supplement Report FDA Received Date05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
412-02-056 LOT 004C1084A; 412-02-056 LOT 004C1084A
Outcome Attributed to Adverse Event Required Intervention;
-
-