ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1140-080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.The reported rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.It is possible that during advancement the balloon became compromised and/or damaged against the left iliac vein or other devices used in the procedure resulting in the reported balloon rupture; however, this could not be confirmed.There is no indication of a product quality issue with respect to manufacture, design or labeling.The additional armada device referenced is being filed under a separate medwatch report number.Na.
 
Event or Problem Description
It was reported that the procedure was to treat a left iliac vein.Post dilatation was to be performed with an armada 35 balloon.The device was advanced to the lesion without resistance and inflated once and ruptured before hitting nominal pressure.Another armada 35 balloon was then advanced and also ruptured before hitting nominal pressure.A third balloon was used successfully to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11562533
Report Number2024168-2021-02331
Device Sequence Number11799905
Product Code LIT
UDI-Device Identifier08717648155048
UDI-Public08717648155048
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberB1140-080
Device Catalogue NumberB1140-080
Device Lot Number01118G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2021
Initial Report FDA Received Date03/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-