• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. CPS SHORT ANCHOR PLUG 12MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. CPS SHORT ANCHOR PLUG 12MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Fracture (1260); Material Twisted/Bent (2981)
Health Effect - Clinical Codes Failure of Implant (1924); Cancer (3262)
Date of Event 03/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical devices: 150465 - oss 5cm diaphyseal segment - 384970; 150354 - oss 7cm segmental femoral rt - 587020; 150493 - oss reinforced yoke - 602080; 150480 - oss axle - 516460; 150477 - oss poly femoral bushings - 790160; 150476 - oss poly tibial bushing - 967230; 150410 - oss tibial poly bearing 12mm - 050900; 150478 - oss poly lock pin - 090530.The customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an oss compress approximately 6 months ago.The surgical team was informed the patient had stopped his chemotherapy, however, at the patient's 2 month post-op appointment, it was noticed that rod from the anchor plug that connects to the spindle had bent and there was no new bone growth.The patient had continued his chemotherapy and the bar fractured and subsequently underwent a revision 2 months later.The surgeon cut just above the anchor plug in order to preserve the pins and bushings and also replaced the femoral component and diaphyseal segment.The tibial component was not revised.The surgical technique was utilized.There was no surgical delay.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6 visual examination of the provided pictures identified a fracture of the anchor plug from the spindle.The dhr was reviewed and no discrepancies relevant to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: fracture of the right knee distal femoral replacement implant.A definitive root cause cannot be determined.Per package insert, fracture is a known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CPS SHORT ANCHOR PLUG 12MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11562603
Report Number0001825034-2021-00861
Device Sequence Number2954998
Product Code KWA
UDI-Device Identifier00880304461703
UDI-Public(01)00880304461703(17)291111(10)037980
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K062998
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number178554
Device Lot Number037980
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/05/2021
Supplement Date Received by Manufacturer06/03/2021
Initial Report FDA Received Date03/24/2021
Supplement Report FDA Received Date06/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age16 YR
Patient Weight81
-
-