| Model Number |
R SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including shock testing and functional testing without duplicating the report.The device was recertified and returned to the customer.Review of the device activity logs showed no evidence of any critical errors that would contribute to the report.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed an "attach pads" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|