This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h3, h6, h10 complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows signs of the outer carton boxes as well as the sterile cavities were damaged.The carton box also exhibited scuffs, crushed corners and edges, and dents.Both the outer and inner sterile cavities are vertically cracked from top to bottom.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue is attributed to transit damage.This event is being reviewed through the capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|