| Catalog Number |
0057621 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was returned for evaluation.Visual evaluation confirms the reported event of a brown colored foreign debris inside of the sealed pouch; however, the origin of the foreign matter could not be determined.As the foreign matter was present in the package prior to sealing, this presented at the time of manufacturing and is confirmed for a manufacturing related condition.Review of manufacturing records indicate product was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in may, 2020.
|
| |
|
Event or Problem Description
|
|
It was reported by the user facility contact that foreign material was noted prior to opening the package of the bard/davol simpulse plus irrigator on (b)(6) 2021.There was no reported patient involvement.
|
| |
|
Search Alerts/Recalls
|