• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC N/A; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC N/A; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734403
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the tool revealed that the returned probe is twisted at the tip and has impact marks at the back end causing fit issues.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information from a manufacturer representative regarding a navigation system outside of a procedure.It was reported that there were several damaged instruments.The cause of the damages are unknown.
 
Event or Problem Description
Medtronic received information from a manufacturer representative regarding a navigation system outside of a procedure.It was reported that there were several damaged instruments.There was a twisted tip, and a piece broken off.The ratcheting egg handle had the top missing off of handle.A navlock was not verify.
 
Additional Manufacturer Narrative
Aware date should have been march 22, 2021.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11564358
Report Number1723170-2021-00787
Device Sequence Number14577138
Product Code OLO
UDI-Device Identifier00613994724014
UDI-Public00613994724014
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734403
Device Catalogue Number9734403
Device Lot Number180418
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2021
Supplement Date Received by Manufacturer04/18/2021
Initial Report FDA Received Date03/24/2021
Supplement Report FDA Received Date04/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-