| Model Number |
1012448-12 |
| Medical Device Problem Codes |
Material Separation (1562); Failure to Advance (2524)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified, heavily tortuous, de novo lesion in the left anterior descending (lad) artery.After stent deployment, a 2.75x12mm nc trek balloon dilatation catheter (bdc) was advanced for post-dilatation but could not cross the lesion and the shaft of the bdc broke in 2 pieces.The device was removed without reported issue.There were no adverse patient effects and no clinically significant delay in the procedure.Another nc trek bdc was used to successfully complete the procedure.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initial filed report, additional information was reported.The shaft separated outside the patient anatomy.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional and functional inspections were performed on the returned device.The reported separation was confirmed.The reported failure to advance could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
| |
|