| Model Number |
2760 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that patients have allegedly sustained burn injuries as a result of the device overheating.The djo investigation is ongoing and a supplemental will be filed upon completion.
|
| |
|
Event or Problem Description
|
|
It was reported that patients have allegedly sustained burn injuries as a result of the device overheating.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that , once applicator is attached the ultrasound wand is not working.Unit keeps on overheating.Returned product inspected and complaint confirmed; no ultrasound output.Weak stim output on ch1 microcurrent.Faulty ultrasound probe's control board.Defective ch1 stim board.Replace pcbs.Calibration required.Calibrate probe.Clean up unit thoroughly inside and out.Software upgrade.Run full functional test.
|
| |
|