MAUDE Adverse Event Report: DJO LLC INTELECT LEGEND XT COMBOUS STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DJO LLC INTELECT LEGEND XT COMBOUS STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT Back to Search Results
Model Number 2760
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Burn(s) (1757); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that patients have allegedly sustained burn injuries as a result of the device overheating.The djo investigation is ongoing and a supplemental will be filed upon completion.
 
Event or Problem Description
It was reported that patients have allegedly sustained burn injuries as a result of the device overheating.
 
Additional Manufacturer Narrative
It was reported that , once applicator is attached the ultrasound wand is not working.Unit keeps on overheating.Returned product inspected and complaint confirmed; no ultrasound output.Weak stim output on ch1 microcurrent.Faulty ultrasound probe's control board.Defective ch1 stim board.Replace pcbs.Calibration required.Calibrate probe.Clean up unit thoroughly inside and out.Software upgrade.Run full functional test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELECT LEGEND XT COMBOUS STD
Common Device Name
STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Manufacturer (Section D)
DJO LLC
5919 sea otter place
ste 200
carlsbad CA 92010
Manufacturer (Section G)
DJ ORTHOPEDICS DE MEXICO, S.A. DE C.V.
carretera libre tijuana tecate
20230 submetropoli el florido
tijuana, mexico 22244
MX   22244
Manufacturer Contact
jim pomeroy
5919 sea otter place
ste 200
carlsbad, CA 92010
5128346301
MDR Report Key11567525
Report Number9616086-2021-00005
Device Sequence Number3279977
Product Code IMG
UDI-Device Identifier00888912292740
UDI-Public00888912292740
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K031077
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2760
Device Catalogue NumberN/A
Device Lot NumberN/A
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2021
Supplement Date Received by Manufacturer07/06/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date07/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-