MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE JEJUNAL FEEDING TUBE; TUBE, FEEDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE JEJUNAL FEEDING TUBE; TUBE, FEEDING Back to Search Results
Model Number M00566340
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive three-port through the peg jejunal tube was used during a jejunal tube placement procedure performed on (b)(6) 2021.During the procedure, when the jejunal tube was inserted via an endovive safety peg tube, the plastic tip detached inside the patient.The detached portion was retrieved endoscopically using forceps.Also, it was noticed that the jejunal tube was difficult to advance into the small intestine.They then decided to discontinue the placement.The procedure was completed with a non bsc device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE JEJUNAL FEEDING TUBE
Common Device Name
TUBE, FEEDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11567546
Report Number3005099803-2021-01037
Device Sequence Number11799946
Product Code KNT
UDI-Device Identifier08714729789666
UDI-Public08714729789666
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K091340
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2022
Device Model NumberM00566340
Device Catalogue Number6634
Device Lot Number0026097948
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2021
Initial Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-