Model Number 1217-12-052
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Muscular Rigidity (1968); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of pinnacle insert.Insert was dislocated from shell with obvious wear.Likely cause impingement.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the affected side was the left hip.It was possible that the liner did not seat properly in the original surgery or that could be due to patient's stiff spine.Liner and head were revised.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Visual examination of the provided x-ray images confirmed the reported event.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : no anomalies or deviations were found during the review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR HA ACET CUP 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11568266
Report Number1818910-2021-06076
Device Sequence Number17009383
Product Code LPH
UDI-Device Identifier10603295009160
UDI-Public10603295009160
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K031495
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-12-052
Device Catalogue Number121712052
Device Lot Number255550
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/15/2021
Supplement Date Received by Manufacturer03/15/2021
05/31/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date04/07/2021
06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
C STEM VOID CENTALIZER 12 DM.; C-STEM SZ3 HIGH OFFSET.; ELITE MODULAR HD 32MM +3.; HARDINGE CEMENT RESTRICTOR.; PINN MAR +4 10D 32IDX52OD.; PINN SECTOR HA ACET CUP 52MM.; C STEM VOID CENTALIZER 12 DM; C-STEM SZ3 HIGH OFFSET; ELITE MODULAR HD 32MM +3; HARDINGE CEMENT RESTRICTOR; PINN MAR +4 10D 32IDX52OD; PINN SECTOR HA ACET CUP 52MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight99
-
-