| Model Number |
FM-410270 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and exchange of textured devices due to patient's concern with product.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d.9, h.3, h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and exchange of textured devices due to patient's concern with product.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and exchange of textured devices due to patient's concern with product.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified yellow particles material was found on the shell and deformation.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.Additional, changed, and/or corrected data: h.3, h.6.
|
| |
|