• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P FOUNDATION KNEE; FEMUR, POROUS COATED, W/NS BEADS, SIZE4, LT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P FOUNDATION KNEE; FEMUR, POROUS COATED, W/NS BEADS, SIZE4, LT Back to Search Results
Model Number 211-02-104
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - loose femoral component,replaced femur and insert.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 6 years and 9 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a ncmr #14918 associated with the main part #211-02-104 femur, porous coated, w/ns beads, size4, lt which documents that out of 9 parts in the lot 1 part was rejected due to contamination in porous coating.Later the rejected part was reworked and accepted after proper justification.All items in the lot have met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this compliant was a revision surgery due to loosening.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
FEMUR, POROUS COATED, W/NS BEADS, SIZE4, LT
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key11568851
Report Number1644408-2021-00218
Device Sequence Number17195004
Product Code JWH
UDI-Device Identifier00888912110020
UDI-Public(01)00888912110020
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923277
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2016
Device Model Number211-02-104
Device Catalogue Number211-02-104
Device Lot Number799A1008
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer04/21/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
385-11-506 LOT 59603927; 385-11-506 LOT 59603927
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-