Model Number 370-13-510
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - revised poly due to failure.The poly insert broke out of locking mechanism.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as to failure of a broken poly insert.The previous surgery and the surgery detailed in this event occurred 9 years and 9 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to failure of a broken poly insert.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
INSERT, CONSTRAINED, P.S., SIZE10 13MM CM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin, tx 78758-5445
MDR Report Key11568857
Report Number1644408-2021-00182
Device Sequence Number14504592
Product Code JWH
UDI-Device Identifier00888912116718
UDI-Public(01)00888912116718
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K933539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2016
Device Model Number370-13-510
Device Catalogue Number370-13-510
Device Lot Number59601115
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date05/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-