• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTALINK PLUS; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTALINK PLUS; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 29720
Medical Device Problem Codes Entrapment of Device (1212); Intermittent Loss of Power (4016)
Health Effect - Clinical Codes Pain (1994); Insufficient Information (4580)
Date of Event 03/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that burr stuck in lesion and open surgery was performed.The target lesion was located in the moderately calcified right posterior tibial artery.A 1.50mm peripheral rotalink plus was selected for use.During procedure, the burr was advanced however it stalled and the patient was in pain.The device was then removed but it only came back at about 3-4cm before getting stuck.Attempts were made to remove the device by using dynaglide, manually pulling back.They even cut the burr off the advancer and used 6fr guide catheter to dislodge the burr from the vessel wall but was unsuccessful.The procedure was then stopped, patient was transferred to the nearby hospital and had an open surgery.The burr was removed and there were perforation seen as it was surrounded by the intimal wall and the patient was doing well post operation.No further patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIPHERAL ROTALINK PLUS
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11569103
Report Number2134265-2021-03739
Device Sequence Number4455401
Product Code MCW
UDI-Device Identifier08714729838906
UDI-Public08714729838906
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K121774
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/24/2022
Device Model Number29720
Device Catalogue Number29720
Device Lot Number0026414410
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2021
Initial Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-