| Model Number |
JR-STE25-2.0-6 |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) for lot 75bj1305 was reviewed.No anomalies were noted on the lot release paperwork.The incident unit was returned for evaluation.The evaluation confirmed a broken firing lever, cracked cartridge housing, and leading cam side staple protrusion.These are indicative of reload jamming.The incidence of reload jamming is rare.
|
| |
|
Event or Problem Description
|
|
It was reported that during a thoracoscopy procedure, the handle of the stapler broke apart while the surgeon was compressing tissue, to staple the lung.The anvil was stuck in a closed position and the surgeon had to manually pry open the jaws.The procedure was completed using clips to stop the bleeding.There were no reported additional patient impacts.
|
| |
|