DEPUY MITEK LLC US 6X12 MILAGRO INTERFRENCE SCREW; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE Back to Search Results
Model Number 231813
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the sales rep in (b)(6) that during a knee arthroscopy procedure on (b)(6) 2021, it was observed that the tip on the screw device broke while it was being inserted into the bone without pressure.The fragments were easily removed.Another like device was used to complete the procedure with a delay of two minutes.The acl was stable even while the screw was broken.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information received, it was reported via complaint submission tool that during a knee arthroscopy while doctor was screwing without pressure a 6x12mm milagro interference screw into the bone, the tip of the screw has broken.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number:7l74291, and no non-conformances were identified.This complaint cannot be confirmed.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6X12 MILAGRO INTERFRENCE SCREW
Common Device Name
ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key11570112
Report Number1221934-2021-00990
Device Sequence Number17024380
Product Code HWC
UDI-Device Identifier10886705021956
UDI-Public10886705021956
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K143660
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number231813
Device Catalogue Number231813
Device Lot Number7L74291
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2021
Supplement Date Received by Manufacturer04/23/2021
04/27/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date04/26/2021
04/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-