• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME, LARGE; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME, LARGE; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number CRANI-L-G1
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported by (b)(6) that during service and evaluation, it was determined that the craniotome device exhibited heat.It was further determined that the device failed pretest for temperature verification.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT CRANIOTOME, LARGE
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key11572080
Report Number1045834-2021-00473
Device Sequence Number19732863
Product Code HBC
UDI-Device Identifier00845384016427
UDI-Public00845384016427
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCRANI-L-G1
Device Catalogue NumberCRANI-L-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2021
Initial Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-