| Model Number |
367859 |
| Medical Device Problem Codes |
Short Fill (1575); Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tubes there was under-fill or low draw of a tube with blood.This event occurred 20 times.The following information was provided by the initial reporter.The customer stated: there was "vacuum loss and low volume draw.".
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tubes there was under-fill or low draw of a tube with blood.This event occurred 20 times.The following information was provided by the initial reporter.The customer stated: there was "vacuum loss and low volume draw.".
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by draw testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identifiy a root cause for the indicated failure mode.
|
| |
|
Search Alerts/Recalls
|