• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Catalog Number 823162
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
2 of 2 reports.Same procedure, different products.Other mfg report number: 3013886523-2021-00130 a facility reported a damaged package of a hakim valve.During procedure, the nurse found that the package of the valve was damaged.Therefore, the physician stop implanting it and changed to a new hakim valve and the procedure was completed.A 15 minute delay was noted with no adverse consequence to the patient.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields:  d4, d9, g3, g6, h2, h3, h4, h6, h10.The valve was returned for evaluation.Device history record (dhr) - product code 82-3162 with lot 3214676, conformed to the specifications when released to stock.Failure analysis - it was not possible to investigate as only the valve was returned in a plastic tub no packaging was returned for investigation.As no defect was reported on the valve this was not investigated.No root cause could not be determined as the no packaging was returned for investigation.The possible root cause for the packaging issue reported by the customer could be due to handling/storage/transport.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Type of Device
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
MDR Report Key11572343
MDR Text Key243861693
Report Number3013886523-2021-00131
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K041296
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Catalogue Number823162
Device Lot Number3214676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Date Manufacturer Received05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-