• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII SLT 3 MOD TIB CUTBLCK RT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII SLT 3 MOD TIB CUTBLCK RT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Model Number 71440205
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during setup or inspection it was noticed that these instruments showed an overabundance of bio-film that has built up on the instruments over the years.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is heavily contaminated.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommended cleaning method given in our cleaning and sterilization brochure ¿recommendations for decontamination and sterilization of smith & nephew orthopedic devices.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SLT 3 MOD TIB CUTBLCK RT
Common Device Name
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11572448
Report Number1020279-2021-02386
Device Sequence Number11668134
Product Code MBH
UDI-Device Identifier03596010460424
UDI-Public03596010460424
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/22/2014
Device Model Number71440205
Device Catalogue Number71440205
Device Lot Number04JM20203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2021
Initial Date Received by Manufacturer 03/03/2021
Supplement Date Received by Manufacturer07/14/2021
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date07/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-