| Model Number |
PM3210 |
| Medical Device Problem Code |
Capturing Problem (2891)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event of high pacing threshold was not confirmed.The pacer was received in normal working conditions with battery voltage at eol.Electrical and mechanical analysis performed, including capture test under nominal settings indicated normal device functionality.The implant duration exceeds the projected longevity based on average monthly current indicating normal battery depletion.Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented for replacement of the pulse generator due to high pacing threshold.The device was explanted and replaced.There were no adverse patient consequences reported.
|
| |
|