• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION EEG-1200A; NEUROFAX ELECTROENCEPHALOGRAPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION EEG-1200A; NEUROFAX ELECTROENCEPHALOGRAPH Back to Search Results
Model Number EEG-1200A
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that their computer turning off unexpectedly during a surgical case.The customer also reported that they were able to recover with minimal data loss and continue recording, but only after having a tech go into the room and turn the recorder back on.No harm or injury was reported.Nihon kohden continues to investigate the reported event.Additional model information: device available for evaluation: the following non-nkc device was used in conjunction with the eeg-1200a: model: a/dell-7050sff s/n: (b)(4).
 
Event or Problem Description
The customer reported that their computer turning off unexpectedly during a surgical case.The customer also reported that they were able to recover with minimal data loss and continue recording, but only after having a tech go into the room and turn the recorder back on.
 
Additional Manufacturer Narrative
Details of complaint: the customer reported that the computer was turning off unexpectedly during a surgical case.They were able to recover with minimal data loss and continue recording, but only after a technician went into the room to turn the recorder back on.No harm or injury was reported.Investigation summary: a root cause could not be determined from the logs provided.There can be several possible causes for a pc suddenly turning off.1) overheating: if the computer's internal components, such as the cpu or graphics card, get too hot, it can trigger a safety mechanism that shuts down the system to prevent damage.Overheating can be caused by insufficient cooling, blocked vents, a malfunctioning fan, or dried-up thermal paste between the cpu and its heatsink.2) power supply issues: a faulty or insufficient power supply unit (psu) may cause the computer to shut down abruptly.If the psu is unable to deliver enough power to the components, the system might turn off to protect itself.3) hardware failure: a malfunctioning hardware component, such as a faulty motherboard, ram module, or hard drive, can cause sudden shutdowns.These failures might be due to age, manufacturing defects, or physical damage.4) software or driver issues: faulty or incompatible software or drivers can cause system instability and unexpected shutdowns.This can occur after installing new software, updating drivers, or experiencing conflicts between different software components.5) electrical issues: problems with the electrical supply, such as power surges or fluctuations, can cause the computer to shut down abruptly.This can happen if the psu or other components are not adequately protected by surge protectors or uninterruptible power supplies (ups).6) virus or malware infections: some malicious software can cause system instability or trigger shutdowns.Running up-to-date antivirus and antimalware scans can help identify and remove any potential threats.A serial number review of the reported device does not reveal additional related complaints.A complaint history review of the customer's account does not reveal trends for similar complaints.The following fields contains no information (ni), as attempts to obtain information were made, but not provided.A2 - a6 b6 b7 attempt #1 03/04/2021 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt #2 03/11/2021 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt #3 03/25/2021 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Additional model information: d10: the following non-nkc device was used in conjunction with the eeg-1200a: model: a/dell-7050sff s/n: (b)(6).
 
Event or Problem Description
The customer reported that the computer was turning off unexpectedly during a surgical case.They were able to recover with minimal data loss and continue recording, but only after a technician went into the room to turn the recorder back on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEG-1200A
Common Device Name
NEUROFAX ELECTROENCEPHALOGRAPH
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key11572948
Report Number8030229-2021-00174
Device Sequence Number11828707
Product Code OLT
UDI-Device Identifier04931921110904
UDI-Public4931921110904
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K080546
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEEG-1200A
Device Catalogue NumberEEG-1200A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2021
Supplement Date Received by Manufacturer05/22/2023
Initial Report FDA Received Date03/25/2021
Supplement Report FDA Received Date05/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A/DELL-7050SFF-W10; COMPUTER: A/DELL-7050SFF-W10
Patient SexUnknown
-
-