• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 7; PROSTH, HIP, SEMI-CONST, MET/CERAM/CERAM/MET, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 7; PROSTH, HIP, SEMI-CONST, MET/CERAM/CERAM/MET, CEMENTED OR UNCEMENTED Back to Search Results
Model Number 71357107
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery took place on (b)(6) 2021 due to an infection.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, a revision was performed due to the patient developing symptoms of infection approximately 2 weeks post implantation.It was communicated that the requested documentation was not available for inclusion in the medical investigation.The clinical root cause of the revision was the reported infection; however, the source of infection is unknown and could not be identified based on the information provided.The patient impact beyond the reported ¿symptoms of infection¿ and the revision could not be determined.No further medical assessment is warranted at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY HO POR PL HA SZ 7
Common Device Name
PROSTH, HIP, SEMI-CONST, MET/CERAM/CERAM/MET, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11573373
Report Number1020279-2021-02404
Device Sequence Number4266271
Product Code MRA
UDI-Device Identifier03596010559623
UDI-Public03596010559623
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P030022
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71357107
Device Catalogue Number71357107
Device Lot Number20HM07997
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/02/2021
Supplement Date Received by Manufacturer08/02/2021
Initial Report FDA Received Date03/26/2021
Supplement Report FDA Received Date08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-