• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SCX-470
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 10/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Event or Problem Description
Healthcare professional reported right side capsular contracture, baker grade iii.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified fold creases.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
 
Event or Problem Description
Healthcare professional reported right side capsular contracture, baker grade iii.Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key11573995
Report Number9617229-2021-03938
Device Sequence Number14332989
Product Code FTR
UDI-Device Identifier10888628030329
UDI-Public10888628030329
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/07/2023
Device Model NumberSCX-470
Device Catalogue NumberSCX-470
Device Lot Number3249299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Initial Date Received by Manufacturer 02/26/2021
Supplement Date Received by Manufacturer08/20/2021
Initial Report FDA Received Date03/26/2021
Supplement Report FDA Received Date09/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
-
-