STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
6260-9-244 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding corrosion involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to stem failure.Intraoperatively, trunnion wear, metallosis, head toggling on trunnion, black tissue, and a broken insert were noted.A 44 mm +4 v40 head, tmzf plus stem, and 0° 44f insert were revised.Rep confirmed that aside from provided primary usage sheet, no further information would be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|