| Model Number |
96-1684 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during the revision knee surgery the 34mm femoral sleeve broach and 14 x 115 stem trial got stuck and was hard to remove.During that time the revision femoral broach handle came off and landed on the floor.We had to get it re-sterilized.While waiting the surgeon started trying to remove with vice grips and damaged the stem trial and sleeve broach.When we did get the handle back it would not stay attached to the broach so i guess it was damaged as well.All three instruments need replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: examination revealed the broach was return separated from the stem trial, thus the jammed/seized condition could not be confirmed.Function evaluation of the two mating devices found the stem trial would not assemble with the broach without the use of excessive force due to severe damage/deformation sustained to the attachment end of the stem trial.No damage to the attachment end of the broach was identified.Examination of the broach did find the cutting edges of the broach are dull and significantly deformed.The investigation did not establish a need for corrective action.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
| |
|