• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752182
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that aspiration could not be performed during ultrasound.The procedure was completed after replacing the product with another one.There's no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and functionally tested.No defects were observed.Both irrigation and aspiration luers were intact.The sample primed and tuned with a handpiece successfully.The sample could be recognized and the service data could be retrieved from the console.Aspiration, max vacuum, and irrigation were tested at appropriate settings and passed testing.The sample passed functionality testing.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11575779
MDR Text Key242910631
Report Number1644019-2021-00209
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657521821
UDI-Public00380657521821
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number8065752182
Device Lot Number2414607H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Date Manufacturer Received06/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
-
-