AMO PUERTO RICO MFG. INC. TECNIS CL IOL; INTRAOCULAR LENS Back to Search Results
Model Number Z9002
Medical Device Problem Codes Difficult to Insert (1316); Inaccurate Delivery (2339)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was fully inserted into patient's right eye, however, was removed and replaced in the same procedure as the lens flipped and was not in correct position.There was incision enlargement and another lens of same model and same diopter power, with serial number (b)(4) was used as replacement.The patient is doing great post operation.No further information is available.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Returned to manufacturer on: apr 15, 2021.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during removal and replacement.Haptic damage (bent) was also observed but can be attributed to handling and cannot be confirmed to be related to manufacturing.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.The search revealed three additional investigation request (ir) for this production order number has been received, however, no product malfunction or deficiency could be identified and no escalations were required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11575817
Report Number2648035-2021-07443
Device Sequence Number4352046
Product Code HQL
UDI-Device Identifier05050474530751
UDI-Public(01)05050474530751(17)241002
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/02/2024
Device Model NumberZ9002
Device Catalogue NumberZ900200230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/26/2021
Supplement Date Received by Manufacturer04/25/2021
Initial Report FDA Received Date03/26/2021
Supplement Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: PSCST30, LOT NUMBER: CH06292.
Outcome Attributed to Adverse Event Required Intervention;
-
-