Model Number N/A
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
This wire was used to navigate a tortuous distal pt.Dr.(b)(6) was attempting to navigate the wire around a tight angle and was unsuccessful.When he removed the wire to use a different wire he noticed that the tip had fractured and the polymer jacket had separated and the inner core of the wire was exposed.Unfortunately another mongo es pre-shaped wire was pulled from the same box with the same lot# and that wire fractured in the same way at the same point of the angled, distal pt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIUS MONGO ES
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ASAHI INTECC
3-100 akatsuki-cho
seto, aichi 489-0 071
JA  489-0071
MDR Report Key11576149
Report Number3004718255-2021-00208
Device Sequence Number16915586
Product Code DQX
UDI-Device Identifier14547327123751
UDI-Public(01)14547327123751(17)230331(10)200424A081
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/12/2021
Report Date (Section F) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberN/A
Device Catalogue NumberAP14R324P
Device Lot Number200424A081
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/12/2021
Device Age10 MO
Event Location Hospital
Date Report to Manufacturer03/26/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA03/26/2021
-
-