Catalog Number 8065752450
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the probe tip was loose, moving freely during surgery.The surgery began and core vit was performed with the probe inserted in the eye, while the foot switch actuated.The tip of the probe then felt gradually stretching out, and was rapidly removed from the eye.Checking the probe tip found the tip free to move probably due to poor fixation.The surgery was completed after replacing the product with another one.There's no patient harm.
 
Manufacturer Narrative
The returned sample was visually inspected and found to be non-conforming with the clear/white foreign material on and around the port and orange/brown foreign material on the welded cap.The sample was then functionally tested for actuation and was found to be non-conforming with the needle stiffener assembly pulling out of the needle holder.The probe sample was disassembled and the components inspected.Nominal wear was observed on the inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks were observed at the cutting edge of the inner cutter.The product was processed and released according to the product¿s acceptance criteria.The evaluation confirms the reported issue of loose probe tip (needle).The root cause for the component detachment is the adhesive bond failure which caused the needle assembly to detach from the rest of the probe assembly.An internal investigation was completed and improvements to the adhesive bond have been identified.A photo of the returned sample has been issued within the learning management system for review with the applicable production personnel to raise awareness of this issue.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed for this reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11576632
MDR Text Key242913277
Report Number1644019-2021-00211
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657524501
UDI-Public380657524501
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number8065752450
Device Lot Number2412426H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2021
Date Manufacturer Received06/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-