Catalog Number 8065752450 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that the probe tip was loose, moving freely during surgery.The surgery began and core vit was performed with the probe inserted in the eye, while the foot switch actuated.The tip of the probe then felt gradually stretching out, and was rapidly removed from the eye.Checking the probe tip found the tip free to move probably due to poor fixation.The surgery was completed after replacing the product with another one.There's no patient harm.
|
|
Manufacturer Narrative
|
The returned sample was visually inspected and found to be non-conforming with the clear/white foreign material on and around the port and orange/brown foreign material on the welded cap.The sample was then functionally tested for actuation and was found to be non-conforming with the needle stiffener assembly pulling out of the needle holder.The probe sample was disassembled and the components inspected.Nominal wear was observed on the inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks were observed at the cutting edge of the inner cutter.The product was processed and released according to the product¿s acceptance criteria.The evaluation confirms the reported issue of loose probe tip (needle).The root cause for the component detachment is the adhesive bond failure which caused the needle assembly to detach from the rest of the probe assembly.An internal investigation was completed and improvements to the adhesive bond have been identified.A photo of the returned sample has been issued within the learning management system for review with the applicable production personnel to raise awareness of this issue.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed for this reported event.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|