Catalog Number 0998-00-0800-53
Device Problems Use of Device Problem (1670); Application Program Version or Upgrade Problem (2881)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Manufacturer Narrative
The customer has not requested getinge to evaluate the iabp in connection with this event.Additional information has been requested, and we will report accordingly if it becomes available.Not returned to manufacturer.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had an error code 14.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
 
Manufacturer Narrative
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (apr 2019 through mar 2021) was reviewed.There were no triggers identified for the review period.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to the site, evaluated the iabp and was unable to reproduce the reported issue.However, the fse was able to verify the alarms in the iabp¿s fault log.The fse found that the iabp unit was out of calibration, so the pressure transducers was calibrated to resolve problem.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had an error code 14.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11577334
MDR Text Key243556780
Report Number2249723-2021-00649
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 06/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/27/2021
Patient Sequence Number1
-
-