SMITH & NEPHEW, INC. TWINFIX ULTRA TI 5.5 W/3 UB; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
72202897 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy using a twinfix ultra, the inserter broke when entering the bone.The pieces were retrieved with tweezers.The original bone hole was expanded to use the backup.The procedure was completed with a 30 minutes or less delay using a back-up device.No further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the subject device and the reported incident.An analysis of the customer provided image revealed that at least three insertor tips had become detached from the devices, and one appeared to be loose.One of the detached tips appeared to be bent.Only one anchor was shown, and it did not appear to have any issues.No debris was shown.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the insertor assembly found that the weld strength is specified and must be inspected for deformities, cracks, and gaps.The complaint was confirmed.Factors that could have contributed to the event include excess force or torque on the device or improper preparation of the insertion site.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|
|
|