Catalog Number 23-2005
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during arthroscopy procedure, outside the patient, the accu-pass direct crescent xl tip is chipped but still on.The procedure was finished with a smith and nephew backup device.Surgical delay less than or equal to 30 mins.Patient injuries were not reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.The tip of the device is detached.A functional evaluation could not be conducted due to detached tip.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The complaint was confirmed.Factor(s) that could have contributed to the reported event include: (1) excessive force.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-PASS DIRECT CRESCENT XL
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11580376
Report Number3006524618-2021-00397
Device Sequence Number20045089
Product Code NBH
UDI-Device Identifier00885556724569
UDI-Public885556724569
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2023
Device Catalogue Number23-2005
Device Lot Number2058254
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/11/2021
Supplement Date Received by Manufacturer05/03/2021
Initial Report FDA Received Date03/28/2021
Supplement Report FDA Received Date06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-