H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one hickman s/l catheter and a separate tissue cuff.Visual, microscopic visual and tactile evaluation were performed on the returned device.The investigation is confirmed for the reported failure to bond issue as the tissue cuff was not secured onto the catheter.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|