| Model Number |
IPN028484 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.A device history review could not be conducted since the lot number provided is not a valid lot number.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
There was direct an air-leak with noise.When pushing hard on the connector it is solved.A new sahara with the same lot number was ok.
|
| |
|
Event or Problem Description
|
|
There was direct an air-leak with noise.When pushing hard on the connector it is solved.A new sahara with the same lot number was ok.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A dhr review could not be conducted since the lot number provided is not a valid lot number.Air source (tfm-0060 nlc09684 due date 22-mar-22) was adapted and connected to section of sample received and ats connector was immerse into the water.No bubbles were observed, therefore the ats does not have leaks.Customer complaint cannot be confirmed.No problem found on sample.
|
| |
|