ZIMMER ORTHOPAEDIC MFG. LTD. FEMORAL COMPONENT PRECOAT SIZE E MINUS (E-) RIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Muscle/Tendon Damage (4532)
Date of Event 11/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical product: 00597003502, tibial component pegged precoat for cemented use only size 4 - 63215771, 00597603010, articular surface anterior constrained size yellow, 64466517.The customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2021-00067, 0001822565-2021-00869.
 
Event or Problem Description
It was reported that a patient underwent an initial right total knee arthroplasty.During the initial procedure, it was noted that the mcl had avulsed from the tibia after placement of all of the final implants.The tear was anchored down with sutures and the procedure completed without further complications.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
His follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The dhr was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: after cementing the final components, the surgeon changed the trial poly to the real poly.During which he noted laxity medially with the real implant and noted the mcl had avulsed from the tibia.Therefore, the surgeon placed two suture anchors and two layers of sutures in the mcl and brought it down nicely to the tibia.It was very stable.No other complication was noted.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No change in the previously found root cause.However, during the investigation, we found that the surgeon used incompatible implant assembly.Per nexgen knee profiler, the combination of gender solutions cr-flex minus precoat femoral is not compatible with nexgen cr/cra (ac) articular surface.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL COMPONENT PRECOAT SIZE E MINUS (E-) RIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key11581846
Report Number3007963827-2021-00077
Device Sequence Number19879456
Product Code JWH
UDI-Device Identifier00889024191228
UDI-Public(01)00889024191228(17)290930(10)64545650
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00575001506
Device Lot Number64545650
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/12/2021
Supplement Date Received by Manufacturer04/29/2021
08/10/2021
Initial Report FDA Received Date03/29/2021
Supplement Report FDA Received Date04/29/2021
08/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age50 YR
Patient Weight79
-
-