• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH- AT ARCADIS VARIC GEN 2; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH- AT ARCADIS VARIC GEN 2; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number 10143406
Medical Device Problem Code Operating System Becomes Nonfunctional (2996)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the arcadis varic gen 2 system.During operation of the system, the user reported that the system hung up.During the error it appeared that the system was totally down.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.The investigation was performed considering complaint description, cs reports and system history.No logs were provided; therefore, a deeper evaluation of this event could not be performed due to the lack of information.According to the customer, the system freezes during operation and no further use of the system could be done.The customer received the message "the application has a serious memory management or internal communication problem" that was displayed at the monitor.The customer did a "forced shut down" of the system by using the reset button.After pressing the on button, the pc did not start, and the system could not be used.When the customer service engineer (cse) arrived onsite and switched the system on, the system started but the error message "internal imaging system error occurred" appears at the monitor.Some software activities were not available.The cse reloaded the software, and the system could be completely recovered afterwards.The patient database was deleted in agreement with the customer.No parts were exchanged and no systematic issue was identified.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCADIS VARIC GEN 2
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key11581892
Report Number3004977335-2021-72561
Device Sequence Number11953798
Product Code OXO
UDI-Device Identifier04056869009018
UDI-Public04056869009018
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051133
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10143406
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer05/20/2021
Initial Report FDA Received Date03/29/2021
Supplement Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-